Technical data

CarboVis 70x IQ

8 - 4

ba75950e02

11/2012

Carbon measurement,

1 mm gap width

Nitrite/nitrate

measurement

5 mm gap width

Measuring

location

Measuring mode

Measuring range

Resolution

Inlet

COD spectr., total

0 ... 20000 mg/L

1 mg/L

COD spectr., dissolv

0 ... 12500 mg/L

1 mg/L

TOC spectr., correl.

0 ... 20000 mg/L

1 mg/L

DOC spectr., correl.

0 ... 12500 mg/L

1 mg/L

BOD spectr., correl.

0 ... 8000 mg/L

1 mg/L

SAC 254, total

0 ... 5000 1/m

1 1/m

UVT 254

0 ... 100.0 %

0.1 %

Activation

COD spectr., dissolv

0 ... 12500 mg/L

1 mg/L

DOC spectr., correl.

0 ... 12500 mg/L

1 mg/L

SAC 254, total

0 ... 5000 1/m

1 1/m

UVT 254

0 ... 100.0 %

0.1 %

Outlet

COD spectr., total

0 ... 4000 mg/L

1 mg/L

COD spectr., dissolv

0 ... 4000 mg/L

1 mg/L

TOC spectr., correl.

0 ... 2500 mg/L

1 mg/L

DOC spectr., correl.

0 ... 2500 mg/L

1 mg/L

BOD spectr., correl.

0 ... 2500 mg/L

1 mg/L

SAC 254, total

0 ... 3000 1/m

1 1/m

UVT 254

0 ... 100.0 %

0.1 %

Measuring

location

Measuring mode

Measuring range

Resolution

Outlet

NO3

0.0 ... 250.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 50.00 mg/L 0.01 mg/L

NO2

0.0 ... 100.0 mg/L

0.1 mg/L

NO2-N

0.00 ... 25.00 mg/L 0.01 mg/L