Technical data

8.1

Measurement characteristics

Measuring principle

Spectrophotometric absorption measurement; integrated micropro-

cessor electronics, shielded 2-wire connection for

power and data transmission.

Wavelength range

200 - 720 nm
200 - 390 nm

Measuring gap (optical

layer thickness)

Carbon measurement,

1 mm gap width

CarboVis 701 IQ

1 mm

CarboVis 705 IQ

5 mm

Measuring

location

Measuring mode

Measuring range Resolution

Inlet

COD spectr., total

0 ... 20000 mg/L

1 mg/L

COD spectr., dissolv

0 ... 12500 mg/L

1 mg/L

TOC spectr., correl.

0 ... 20000 mg/L

1 mg/L

DOC spectr., correl.

0 ... 12500 mg/L

1 mg/L

BOD spectr., correl.

0 ... 8000 mg/L

1 mg/L

SAC 254, total

0 ... 5000 1/m

1 1/m

SAC 254, dissolved

0 ... 3000 1/m

1 1/m

UVT 254

0.0 ... 100.0 %

0.1 %

Activation

COD spectr., dissolv

0 ... 12500 mg/L

1 mg/L

DOC spectr., correl.

0 ... 12500 mg/L

1 mg/L

SAC 254, total

0 ... 5000 1/m

1 1/m

SAC 254, dissolved

0 ... 3000 1/m

1 1/m

UVT 254

0.0 ... 100.0 %

0.1 %