7 what to do if, What to do if -1, 7what to do if – YSI CarboVis 705 IQ User Manual

Page 45

Advertising
background image

CarboVis 70x IQ

What to do if...

7 - 1

ba75950e02

11/2012

7

What to do if...

"----" display

(invalid measured value)

Implausible measured

values

Cause

Remedy

– User calibration values

entered incorrectly

*

Sensor check (H2O dest.)

or

*

Zero adjustment

or

*

Sensor check (standard)

defective

– Correct the entry and enter

once again (section 4.3.2)

Repeat

*

Sensor check (H2O dest.)

or

*

Zero adjustment

or

*

Sensor check (standard)

.

Check conditions, especially

cleanliness (section 4.3.3)

Zero adjustment

still errone-

ous after several trials

– Contact WTW

– Unknown

– See log book

Cause

Remedy

– Erroneous user calibration

– Repeat user calibration. Make

sure the samples for compari-

son are representative.

– Substantial change of test

sample (sample matrix)

– Carry out new user calibration

– Measurement windows dirty

– Clean the measurement win-

dows

– Sensor setting wrong

– Check and, if necessary, cor-

rect the

Measuring mode

,

Mea

-

suring location

sensor settings

– Carry out new user calibration

– Optical system maladjusted,

e. g. due to inadmissible

mechanical shock

– Carry out

*

Sensor check (H2O dest.)

or

*

Zero adjustment

or

*

Sensor check (standard)