2 application conditions, 2 application conditions -5 – YSI CarboVis 705 IQ User Manual

Page 53

Advertising
background image

CarboVis 70x IQ

Technical data

8 - 5

ba75950e02

11/2012

Carbon measurement,

5 mm gap width

The quoted measuring ranges are nominal measuring ranges that are

theoretically possible. In practice, real measuring ranges exist that are

given by the limits of photometric determination. The limits are signifi-

cantly influenced by the light scattering due to solids and the absorption

of accompanying substances (sample matrix). Therefore, no fixed

measuring range can be determined beforehand.

8.2

Application conditions

Suitability and areas of

application

Municipal wastewater with a low proportion of industrial wastewater

Measuring medium:

Pressure resistance

Sensor with connected SACIQ (SW) sensor connection cable:

The sensor meets all requirements according to article 3(3) of 97/23/

EC ("pressure equipment directive").

Type of protection

Sensor with connected SACIQ (SW) sensor connection cable:
IP X8, 1 bar (10

5

Pa)

Measuring

location

Measuring mode

Measuring range

Resolution

Outlet

COD spectr., total

0.0 ... 800.0 mg/L 0.1 mg/L

COD spectr., dissolv

0.0 ... 800.0 mg/L 0.1 mg/L

TOC spectr., correl.

0.0 ... 500.0 mg/L 0.1 mg/L

DOC spectr., correl.

0.0 ... 500.0 mg/L 0.1 mg/L

BOD spectr., correl.

0.0 ... 500.0 mg/L 0.1 mg/L

SAC 254, total

0.0 ... 600.0 1/m

1 1/m

UVT 254

0.0 ... 100.0 %

0.1 %

pH value

4 - 12

Flow rate

< 3 m/s

Temperature

0 °C ... + 45 °C

(32 °F ... + 113 °F)

Max. allowed overpressure

105 Pa (1 bar)

Advertising
This manual is related to the following products: