9.1.1 Error messages

Message code

Message text

EA2541

Sensor temperature too high!

* Check process and application

EA3541

Sensor temperature too low!

* Check process and application

EAI541

Zero adjustment erroneous

* Clean the sensor and sensor sleeve repeatedly

* Position the sensor sleeve according to operating manual

* Rinse the measuring chamber with ultrapure water several times

* Fill the measuring chamber with ultrapure water

* Repeat the zero adjustment

EAM3Ax

Meas. range exceeded or undercut

* Check process

EC33Ax

User calibration erroneous,

Check raw value / reference value pairs:

* Set all values within the measuring range

(see operating manual)

* Enter value pairs in ascending order

EI1541

Operational voltage too low

* Check installation and cable lengths, Follow installation instructions

* Power supply module(s) overloaded, add power supply module(s)

* Check terminal and module connections

* Defective components, replace components

EI2541

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power supply module(s) overloaded, add power supply module(s)
* Check terminal and module connections

* Defective components,

replace components

ES2541

Humidity to high in sensor

* Contact service

ES3541

Failure optical measurement: deviation to calibration for reference

channel too high

* Contact service

ES4541

Failure optical measurement: hardware lamp

* Contact service