YSI IQ SensorNet NitraLyt Sensor User Manual

Page 16

Advertising
background image

Commissioning

NitraLyt

®Plus

700 IQ

3 - 4

ba76004e01

01/2012

3.3.2

Measurement electrodes and compensation electrodes

Commissioning

Fig. 3-2

Measurement electrode or compensation electrode with storing aids

In the delivery condition, each electrode is equipped with a watering
cap and a nut that protects the screw-in thread. Prior to installation, first
remove the watering cap and, using the special hexagon key, unscrew
the electrode from the nut. Keep both storing aids in case you want to
store the electrode.

Caution

The membrane of the electrode must not

dry up (follow notes on storage)

be damaged

be brought into contact with grease.

Notes on storage

If you will not use the electrode for a longer period of time, screw the
electrode into the nut as far as it will go. Soak the foam insert in the
watering cap with VARiON

®

/ES-1 standard solution (lower

concentration) and plug the electrode into the watering cap.

Caution

Make sure to use the correct solution for the watering cap
(VARiON

®

/ES -1 standard solution). If you use the watering

solution of the reference electrode instead the function of the
electrode can be seriously damaged.

Watering cap

Electrode

Nut

Sealing ring

Advertising