NitraLyt

®Plus

700 IQ

Overview

1 - 3

ba76004e01

01/2012

Operating modes and

electrode equipment

Due to its modular structure, the NitraLyt

®Plus

700 IQ can be adapted

to various requirements (see following table).

Notes:

The reference electrode has an extra receptacle marked by a recess.
The ion sensitive electrodes can be mounted in the remaining two
receptacles in any order. Empty receptacles have to be closed with the
VARiON

®

BP blind plug.

Calibration free

operation

The NitraLyt

®Plus

700 IQ nitrate sensor is immediately ready to

measure after being equipped with electrodes. For precise
measurements, it is only necessary to adjust the electrodes to the
sample matrix ("matrix adjustment"). In the recommended application
(see section 1.3 R

ECOMMENDED

FIELDS

OF

APPLICATION

), the measuring

characteristics of the electrodes remain stable for their entire service
life. Thus, calibration is no longer required.

Possible changes of the sample matrix can be determined by
occasional comparison measurements (e.g. photometer) and
compensated for by a new matrix adjustment. The sensor does not
have to be taken out of the sample for this.

Shielding of the

NitraLyt

®Plus

700 IQ

The NitraLyt

®Plus

700 IQ nitrate sensor and the corresponding

electrodes in conjunction with the

IQ S

ENSOR

N

ET

system form a

measuring system that is protected to a high degree against low and
high frequency interference as well as against the indirect effects of
lightning strikes.

Operating mode

Electrode equipment

Nitrate measurement

Ref

VARiON

®Plus

NO3

Blind plug

VARiON

®

Ref

Advertising