01/2012

Faulty result of the

matrix adjustment

Faulty result of

calibration

Cause

Remedy

– Error during procedure, e.g.

incorrect lab values

– Check the basic conditions

– Follow the practical notes on

page 4-4 or in the online help

– Recalibrate the electrode

– Electrode not sufficiently

adapted (conditioned) to test
sample

– Switch to mV mode and

observe measured value

– When measured value stable,

repeat matrix adjustment

– Reference electrode or

measurement electrode
defective due to aging (see
section 8.1)

– Replace defective electrode

Cause

Remedy

– Error during procedure, e.g.

contaminated standard
solutions

– Check the basic conditions

– Follow the practical notes on

page 4-10 or in the online help

– Recalibrate the electrode

– Reference electrode or

measurement electrode
defective due to aging (see
section 8.1)

– Replace defective electrode