YSI IQ SensorNet NitraLyt Sensor User Manual

Page 30

Advertising
background image

Matrix adjustment, check and calibration

NitraLyt

®Plus

700 IQ

4 - 6

ba76004e01

01/2012

The history of matrix adjustments contains the following information:

Note
The history of matrix adjustments of the chloride electrode is not output
with the calibration history of the sensor (NitraLyt+). If necessary, you
can create a log book message with the calibration history by means of
the menu item, Calib. history Cl in the setting table. How to do this is
described on page 3-10. The log book message is given in the log book
(message code IC5396).

Date

Date of the matrix adjustment

S

Slope [mV] of the electrode.
The slopes are marked by a star (*) because they were not changed
by the matrix adjustment or reset to the default condition with the
electrode zeroing.

DV

Drift voltage [mV]
0 is displayed after an electrode zeroing.

Ref1

Reference concentration [mg/l]

Ref2

This column is of no importance for the matrix adjustment

Cl-

Measured or entered chloride concentration [mg/l]

P

Procedure (1 = matrix adjustment)

T

Temperature [°C]

R

Evaluation of the matrix adjustment
+ : Matrix adjustment successful.

The sensor measures with the new adjustment data.

? : Matrix adjustment unsuccessful.

Invalid adjustment data was discarded. Measurement is
continued with the last valid values.