YSI IQ SensorNet NitraLyt Sensor User Manual

Page 28

Advertising
background image

Matrix adjustment, check and calibration

NitraLyt

®Plus

700 IQ

4 - 4

ba76004e01

01/2012

Practical instructions

The nitrate concentration has to be determined immediately after
taking the sample as the nitrate content changes very quickly due to
the micro organisms that are present. It is best to take the sample
using a syringe filter for transport to the laboratory or to stabilize it
otherwise. When adding stabilizing solutions, the dilution factor has
to be taken into account.

While determining the reference concentrations in the lab (step 2),
you can use the sensor for control purposes again by switching the
maintenance condition off and thus releasing the linked outputs. The
sensor continues to use the data of the previous matrix adjustment
(or calibration). The reference voltages determined in step 1 will not
be lost. They remain stored until step 3 of the matrix adjustment is
completed. They do not have to be noted and entered again.

Use the Online help if you are unsure of anything during the matrix
adjustment.