2 info messages, 2 info messages -2 – YSI IQ SensorNet NitraLyt Sensor User Manual
Page 60
Indexes
NitraLyt
®Plus
700 IQ
11 - 2
ba76004e01
01/2012
11.1.2 Info messages
EIA523
Incorrect equipment
* for correct electrode equipment see operating manual
ES1523
Component hardware defective
* Contact service
Message code
Message text
Message code
Message text
IC4523
Cl electrode has been successfully calibrated
* For calibration data, see calibration history
IC6396
(this message contains calibration data of the chloride electrode)
IC7396
Sensor could not be calibrated,
Measuring with old calibration values
* Check calibration conditions and calibration standard
* View calibration history
* Service sensor immediately
(see operating manual)
ICA396
Electrode: check successful
ICC396
Cl electrode: check successful
ICD396
Electrode: check unsuccessful
Please follow online help.
ICF396
Cl electrode: check unsuccessful
Please follow online help.
IIA523
(This message is generated when the electrode equipment is changed.
It informs you of the new assignment of the electrode receptacles)