2 mounting the protective hood, Mounting the protective hood -6 – YSI IQ SensorNet NitraLyt Sensor User Manual

Page 18

Advertising
background image

Commissioning

NitraLyt

®Plus

700 IQ

3 - 6

ba76004e01

01/2012

The measurement and compensation electrode can be mounted in
the remaining two receptacles in any order.

During installation make sure the area behind the sealing ring of the
electrode and the receptacle are absolutely dry and clean.

Plug the electrode on the special socket wrench provided and insert
the electrode with the special socket wrench.

Screw until the electrode seats without any gap on the electrode
support. Thus the tightness and electrical contacting are granted.

Note
When mounted, the electrodes can be recognized by the features
described in section 6.3.

3.4.2

Mounting the protective hood

CH cleaning head

(option)

For permanent operation, we recommend to use the CH cleaning head
for compressed air-driven cleaning. It is mounted instead of the
standard protective hood. The compressed air cleaning is started time-
controlled via the IQ S

ENSOR

N

ET

system. Information on the required

components is given in the YSI catalog and on the Internet.

If no CH cleaning head is used the standard protective hood should
always be mounted for measuring. It protects the electrodes from rough
mechanical impact.

Advertising