2 nitralyt+ setting table, Nitralyt+ setting table -10 – YSI IQ SensorNet NitraLyt Sensor User Manual

Page 22

Advertising
background image

Commissioning

NitraLyt

®Plus

700 IQ

3 - 10

ba76004e01

01/2012

3.5.2

NitraLyt+ setting table

Menu item

Selection/values

Explanations

Measuring mode

NO3-N

NO3

mV

Citation form of the mass concentration or
voltage of the electrode.

Measuring range

(Measuring mode: NO3-N)

AutoRange

0.1 ... 100.0 mg/l

1 ... 1000 mg/l

2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.

Measuring range

(Measuring mode: NO3)

AutoRange

0.5 ... 450.0 mg/l

5 ... 4500 mg/l

2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.

Measuring range

(Measuring mode: mV)

-2000 ... 2000 mV

Fixed range

Chloride compens.

automatic / manual
1 ... 1000 mg/l

automatic

(with a chloride electrode mounted):
When a chloride electrode is mounted, the
chloride compensation takes places
automatically only. The value of the
chloride concentration measured at the
time the setting table was opened is
displayed in the next line.

manual

(without any chloride electrode mounted):
After determining the chloride content of
the test solution enter the determined
chloride content manually in the next line.
TThe measured value is accordingly
corrected by the entered chloride content.
Note: Detailed information on the subject
of chloride compensation is given in the
YSIYSI primer, I

ON

SELECTIVE

MEASUREMENT

IN

ONLINE

ANALYSIS

.

Calib. history Cl
(only with Chloride
compens.
automatic)

Do not download

Transmit to log book

Send to log book generates a log book
message with the calibration history of the
chloride electrode. When opening the
setting table again the setting is reset to
Do not download.