Matrix adjustment, check and calibration

4 - 1

ba76004e01

01/2012

4

Matrix adjustment, check and calibration

4.1

General information

Calibration free

operation

The NitraLyt

®Plus

700 IQ nitrate sensor is immediately ready to

measure after being equipped with electrodes. For precise
measurements, it is only necessary to adjust the electrodes to the
sample matrix ("matrix adjustment"). In the recommended application
(see section 1.3 R

ECOMMENDED

FIELDS

OF

APPLICATION

), the measuring

characteristics of the electrodes remain stable for their entire service
life. Thus, calibration is no longer required.

Possible changes of the sample matrix can be determined by
occasional comparison measurements (e.g. photometer) and
compensated for by a new matrix adjustment. The sensor does not
have to be taken out of the sample for this. At the same time, a matrix
adjustment provides information on the state of the electrodes.

Steps

Fig. 4-1

Schematic diagram

Start: press

<C>

Electrode

zeroing

Yes/No

Selection of procedure

Matrix adjustment

Check

Calibration

Procedures for special cases

(see section 4.3)

Calibration free operation

Yes if a new or different electrode is

commissioned