11 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet NitraLyt Sensor User Manual

Page 59: 11 indexes -1, 1 explanation of the messages -1, 1 error messages -1

Advertising
background image

NitraLyt

®Plus

700 IQ

Indexes

11 - 1

ba76004e01

01/2012

11

Indexes

11.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S

ENSOR

N

ET

system for the NitraLyt

®Plus

700 IQ sensor.

Note
Information on

the contents and structure of the log book and

the structure of the message code

is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system

operating manual.

Note
The last three digits of the message code identify the source of the
message:

523 = NitraLyt+700 IQ (armature / component class, ADA adapters)

396 = NitraLyt+ (nitrate sensor)

11.1.1 Error messages

Message code

Message text

EA1396

Meas. range exceeded or undercut
* Check process
* Select other meas. range

EA2523

Sensor temperature too high!
* Check process and application

EA3523

Sensor temperature too low!
* Check process and application

EAN396

Chloride measurement: range exceeded or undercut
* Check process

EC7523

Cl electrode could not be calibrated, nitrate measurement blocked,
nitrate measurement possible with manual Cl compensation and Cl
electrode removed
* Check calib. conditions and standard
* View calibration history
* Service sensor immediately

(see operating manual)