2 status info, 2 status info -3 – YSI IQ SensorNet NitraLyt Sensor User Manual

Page 61

Advertising
background image

NitraLyt

®Plus

700 IQ

Indexes

11 - 3

ba76004e01

01/2012

11.2

Status info

The status info is a piece of coded information about the current state
of a sensor. Each sensor sends this status info to the controller. The
status info of sensors consists of 32 bits, each of which can have the
value 0 or 1.

Status info,

general structure

The bits 0 - 15 are reserved for general information.

The bits 16 - 21 are reserved for internal service information.

You obtain the status info:

via a manual query in the menu, Einstellungen/Settings/Service/List
of all components
(see system operating manual)

via an automated query

– of a superordinate process control (e. g. when connected to the

Profibus)

– of the IQ Data Server (see operating manual of the

IQ S

ENSOR

N

ET

software pack)

Note
The evaluation of the status info, e.g. in the case of an automated
query, has to be made individually for each bit.

NitraLyt

®Plus

700 IQ

Status info

0 1 2 3 4 5 6 7

8 9 10 11 12 13 14 15

1 0 0 0 0 0 0 0

0 0 0 0 0 0 0 0

(general)

0 0 0 0 0 0 0 0

0 0 0 0 0 0 0 0

(internal)

16 17 18 19 20 21 22 23

24 25 26 27 28 29 30 31

Status bit

Explanation

Bit 0

Component hardware defective

Bit 1-31

-

Advertising