MIQ/C6
Installation
3 - 3
ba76033e01
01/2012
General installation
instructions
Observe the following points when attaching connecting wires to the
terminal strip
Shorten all wires to be used to the length required for the installation
Always fit all the ends of the wires with wire end sleeves before
connecting them to the terminal strip
Any wires that are not used and project into the enclosure must be
cut off as closely as possible to the cable gland.
Screw a small cable gland with sealing ring into each remaining free
opening and close it with a blind plug.
3.4
Connections to the current outputs
Materials required
Wire end sleeves, suitable for the connecting wires, with suitable
crimping tool
4 x cable gland with sealing ring (scope of delivery MIQ/C6)
Tools
Cable stripping knife
Wire stripper
Phillips screw driver
Small screw driver