4 connections to the current outputs, Connections to the current outputs -3 – YSI IQ S ENSOR N ET MIQ/C6 User Manual

Page 11

Advertising
background image

MIQ/C6

Installation

3 - 3

ba76033e01

01/2012

General installation

instructions

Observe the following points when attaching connecting wires to the
terminal strip

Shorten all wires to be used to the length required for the installation

Always fit all the ends of the wires with wire end sleeves before
connecting them to the terminal strip

Any wires that are not used and project into the enclosure must be
cut off as closely as possible to the cable gland.

Screw a small cable gland with sealing ring into each remaining free
opening and close it with a blind plug.

3.4

Connections to the current outputs

Materials required

Wire end sleeves, suitable for the connecting wires, with suitable
crimping tool

4 x cable gland with sealing ring (scope of delivery MIQ/C6)

Tools

Cable stripping knife

Wire stripper

Phillips screw driver

Small screw driver