Connecting lines to the terminal strip, Installation miq/c6 3 - 4, Current outputs – YSI IQ S ENSOR N ET MIQ/C6 User Manual

Page 12

Advertising
background image

Installation

MIQ/C6

3 - 4

ba76033e01

01/2012

Connecting lines to the

terminal strip

Fig. 3-1

Terminal strip with the current connections

1

Open the module.

2

Open the dummy screw fitting under the required connection.
Keep the dummy screw fitting for possible later modifications.

3

Screw the cable gland (pos. 1 in Fig. 3-1) with the sealing ring
(pos. 2) into the module housing.

4

Loosen the coupling ring (pos. 3 in Fig. 3-1).

5

Feed the line through the cable gland in the module housing.

6

Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the label located under the
terminal strip.

7

Tighten the coupling ring (pos. 3 in Fig. 3-1).

8

Close the module.

1
3

2

Current outputs

X18

+ REC3 -

+ REC6 -

+ REC2 -

+ REC5 -

SENSORNET 2

SENSORNET 1

+ REC1 -

+ REC4 -

ON

OFF

0/4...20mA

0/4...20mA

0/4...20mA

0/4...20mA

0/4...20mA

0/4...20mA

X17

X16

X15

X14

X13

X12

X11

X10

X8

X9

X7

X6

RED

RED

SN

TERMINA

TO
R

GREEN

SHIELD

GREEN

X5

X4

X3

X2

X1

SHIELD