6 technical data, Technical data -1, Chapter 6 t – YSI IQ S ENSOR N ET MIQ/C6 User Manual

Page 33: Echnical, Data, Ee chapter 6 t, 6technical data

Advertising
background image

MIQ/C6

Technical data

6 - 1

ba76033e01

01/2012

6

Technical data

Note
General technical data on MIQ modules are given in the T

ECHNICAL

DATA

chapter of the IQ S

ENSOR

N

ET

system operating manual.

Electrical data

Instrument safety

Terminal connections

Nominal voltage

Max. 24 VDC via the IQ S

ENSOR

N

ET

(for

details, see the T

ECHNICAL

DATA

chapter of

the IQ S

ENSOR

N

ET

system operating

manual)

Power consumption

3 W

Protective class

II

Overvoltage
category

II

Applicable norms

– EN 61010-1

– UL 3111-1

– CAN/CSA C22.2 No. 1010.1

IQ S

ENSOR

N

ET

connections

2
Additional connectable SENSORNET
terminator (terminating resistor)

Number of current
outputs

6

Terminal type

Screw-type terminal strip, accessible by
opening the lid

Terminal ranges

Solid wires:

0.2 ... 4.0 mm

2

AWG 24 ... 12

Flexible wires:

0.2 ... 2.5 mm

2

Line cross-section of
cables carrying mains
voltage

Europe:

1.5 ... 4.0 mm

2

USA:

AWG 14 ... 12

Cable feeds

4 cable glands M16 x 1.5 on the underside
of the module