1 recorder – YSI IQ S ENSOR N ET MIQ/C6 User Manual

Page 18

Advertising
background image

Settings

MIQ/C6

4 - 6

ba76033e01

01/2012

4.4.1

Recorder

Function

In the Recorder application, the measured values of the linked sensor
are set up as current intensity at the current output. The output of
measured values is laid down in the Recorder type, Start value and End
value
settings.

Settings

8

Carry out the settings for the current output with

d

and

g

.

9

Highlight and confirm Save and quit with

d

and

g

.

The new settings are stored.

Setting

Selection/Values

Explanation

Recorder type

0 to 20 mA or
4 to 20 mA

Start value

within the
measuring range
(sensor-
dependent)

Minimum spacing:
5 % of the measuring
range

End value

Measured variable

Main variable
Adjoining variable

Main variable designates
the actual measured
variable of the sensor
(e.g. pH, oxygen, etc.).
Adjoining variable
designates an additional
measured variable (e.g.
temperature).

Attenuation

0 ... 40 mA/s

Speed of change of the
output current (mA/s) in
the case of an erratic
change of the input
signal.

I -> OFL/UFL

Error

Current values outside
the range between Start
value
and End value are
considerd as errors. The
current output reacts as
specified under Behavior
at error
(see below).

Limitation

The current at the output
is limited to the Start
value
or End value.