Commissioning

3 - 7

ba76000e01

01/2012

3.4.2

Mounting the protective hood

CH cleaning head

(option)

For permanent operation, we recommend to use the CH cleaning head
for compressed air-driven cleaning. It is mounted instead of the
standard protective hood. The compressed air cleaning is started time-
controlled via the IQ S

ENSOR

N

ET

system. Information on the required

components is given in the YSI catalog and on the Internet.

If no CH cleaning head is used the standard protective hood should
always be mounted for measuring. It protects the electrodes from rough
mechanical impact.

Mounting the standard

protective hood

Fig. 3-4

Mounting the standard protective hood.

Cleaning the protective

hood

The coupling ring of the protective hood can be taken apart for cleaning
purposes (see section 6.2 E

XTERIOR

CLEANING

).

1

Loosen the coupling ring (1) of the protective hood.

2

Push the protective hood on the sensor as far as it will go.

3

Tighten the coupling ring of the protective hood.