Commissioning

VARiON

®Plus

700 IQ

3 - 16

ba76000e01

01/2012

3.5.4

Setting table of the VARiON K (potassium sensor)

Note
These sensor settings are only available if in the sensor settings for the
VARiON A the VARiON K setting was set to Active (see section 3.5.2).

The basic settings are taken over from the VARiON A sensor but can
be adjusted separately afterwards.

Menu item

Selection/values

Explanations

Measuring mode

K (mg/l)
mV

Citation form of the mass concentration or
voltage of the electrode.

Measuring range
(Measuring mode: K (mg/l))

1 ... 1000 mg/l

Fixed range

Measuring range
(Measuring mode: mV)

-2000 ... 2000 mV

Fixed range

IQ-LabLink

Not active

(not used)

Temperature mode
(only with
Measuring mode: K (mg/l))

°C

°F

Unit of the measured temperature value
(Celsius, Fahrenheit).

Temp. adjustment
(only with
Measuring mode: K (mg/l))

-1.5 °C ... +1.5 °C

The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:
 Due to the thermal capacity of the

sensor, it is necessary to place it in a
container with at least 2 liters of water.

 Leave the sensor in this container for at

least 15 minutes while stirring
occasionally, then carry out the
adjustment.
If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.

Conc. offset

-10 mg/l ... +10 mg/l

Constant value, which is added to the
measuered value.

Program level

0 ... 2

Program extension (0 = Off = standard set-
ting).

Refer to section 4.2.4 for details about the
program extension.