YSI IQ S ENSOR N ET VARiON Plus 700 IQ User Manual

Page 57

Advertising
background image

VARiON

®Plus

700 IQ

What to do if ...

8 - 3

ba76000e01

01/2012

Measurement provides

jumping, unstable or

drifting values

– Electrode contaminated

– Clean the electrode (see

section 6.2)

– Liquid in the sensor shaft

– Sensor defective, send it back

Cause

Remedy

Cause

Remedy

– Measurement /

compensation electrode:
Electrode membrane not
moistened by measuring
solution, e.g. due to air in
front of the membrane

– Moisten membrane with

deionized water using a wash
bottle. To do so, position the
opening of the wash bottle on
the membrane and splash
against the membrane
vigorously

– Measurement /

compensation electrode:
Air bubble behind the
membrane

– Hold the electrode in a vertical

position with the membrane
down and knock it on the side
with the special socket wrench

– Measurement electrode /

compensation and reference
electrode:
Insufficient electrical contact
in the electrode receptacle

– Check the installation and

contacts of the electrode (gap-
free mounting)

– Check electrode receptacle for

moisture

– If necessary, unscrew the

electrode/blind plug and
thoroughly dry the electrode/
blind plug and receptacle

– Reference electrode dried up

– Replace reference electrode

– Measurement /

compensation electrode or
reference electrode leaky or
damaged

– Replace defective electrode

– Liquid in the sensor shaft

– Sensor defective, send it back

Advertising