2 setting table of the varion a (ammonium sensor), Setting table of the varion a (ammonium sensor) – YSI IQ S ENSOR N ET VARiON Plus 700 IQ User Manual

Page 23

Advertising
background image

VARiON

®Plus

700 IQ

Commissioning

3 - 11

ba76000e01

01/2012

3.5.2

Setting table of the VARiON A (ammonium sensor)

Menu item

Selection/values

Explanations

Measuring mode

NH4-N
NH4
mV

Citation form of the mass concentration or
voltage of the electrode.

Measuring range
(Measuring mode: NH4-N)

AutoRange
0.1 ... 100.0 mg/l
1 ... 1000 mg/l

2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.

Measuring range
(Measuring mode: NH4)

AutoRange
0.1 ... 130.0 mg/l
1 ... 1290 mg/l

2 measuring ranges can be selected.
With AutoRange, the instrument
automatically switches to the suitable
measuring range.

Measuring range
(Measuring mode: mV)

-2000 ... 2000 mV

Fixed range

VARiON K

hide
show

hide (standard setting):
The potassium electrode is not displayed
as an extra sensor in the measured value
display.

show:
If the sensor is equipped with the
potassium electrode, the electrode is
displayed as the extra sensor VARiON K
in the measured value display.
The relevant settings can be done in an
extra setting table (see section 3.5.4).
Note: If the maximum number of sensors
for the system is exceeded by activating
the sensor, the VARiON K cannot be
activated.

Calib. history K
(only with setting VARiON
K
: hide)

Do not download

Transmit to log book

If you select Transmit to log book, a log
book message with the calibration history
of the potassium electrode is generated.
When opening the setting table again the
setting is reset to Do not download.