2 error causes and remedies, Error causes and remedies -2 – YSI IQ S ENSOR N ET VARiON Plus 700 IQ User Manual

Page 56

Advertising
background image

What to do if ...

VARiON

®Plus

700 IQ

8 - 2

ba76000e01

01/2012

8.2

Error causes and remedies

No measured value

display

Measurement provides

implausible measured

values

Cause

Remedy

– Sensor not connected

– Connect the sensor

– Incorrect electrode

equipment

– Correct electrode equipment

– Electrode(s) not at all or

incorrectly recognized by the
system

– Check the installation and

contacts of the electrode (gap-
free mounting)

– Check electrode receptacle for

moisture

– If necessary, unscrew the

electrode/blind plug and
thoroughly dry the electrode/
blind plug and receptacle

– Unknown

– Look for error messages in the

log book

– Liquid in the sensor shaft

– Sensor defective, send it back

Cause

Remedy

– No matrix adjustment or

calibration carried out

– Adjust or calibrate the

electrode

– Adjustment or calibration

error (e.g. incorrect lab
values, contaminated
standard solutions)

– Check the adjustment or

calibration conditions

– Readjust or recalibrate the

electrode

– Manual interfering ions

compensation works with an
unsuitable value

– Determine and enter the

interfering ions compensation
once again, then recalibrate

– Electrode(s) not at all or

incorrectly recognized by the
system

– Check the installation and

contacts of the electrode (gap-
free mounting)

– Check electrode receptacle for

moisture

– If necessary, unscrew the

electrode/blind plug and
thoroughly dry the electrode/
blind plug and receptacle