Dissolved oxygen (D.O.)

MultiLab 4010-3

50

ba76143e01

04/2013

7.2.5 2-point calibration

Calibration in water vapor-saturated air (Cal) and calibration via Comparison
meas.
according to the Winkler method (Comp) are described jointly here.

5.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

6.

Wait for the AutoRead measurement to be completed (audio signal).

When calibrating in water vapor-saturated air (Cal):
The calibration record is displayed and output to the interface.

When calibrating via comparison measurement (Comp):
Set the concentration value with <> <> and accept it with
<ENTER>.
The calibration record is displayed and output to the interface.

7.

Switch to the measured value display with <ENTER>.

1.

Connect the D.O. sensor to the meter.

2.

Check the selected calibration type and selected number of calibration
points in the sensor setting menu and, if necessary, change them.

03.04.2013 08:00

03.04.2013 08:00