9.3.2 Enter Sensor Cap coefficients

Salinity from chan-
nel

▭▭▬

▭▬▭

▬▭▭

Selects the channel (sensor), that pro-
vides the measured temperature
value.

This menu item is only available if an
IDS adapter and two IDS sensors with
temperature sensor are connected.

Sal correction

on
off

Manual salt content correction for con-
centration measurements.

Calibration /
Sensor Cap coeffici
ents

K1
...
K5
KC

Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2 E

NTER

S

ENSOR

C

AP COEFFICIENTS

, page 67

Salinity

0.0 ... 70.0

Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
automatic salt content correction is
switched off and the manual salt con-
tent correction is switched on.

Resolution

0.1
1

Set a high or low resolution.
The setting of the resolution is sepa-
rately stored for each measured
parameter.

Saturation local

on
off

Saturation local is a procedure that
takes the local air pressure into
account for each saturation measure-
ment.
For details, see section 9.3.3
S

ATURATION LOCAL

, page 68.

Reset

-

Resets all sensor settings to the deliv-
ery condition (see section 9.6.1 R

ESET

-

TING

THE

MEASUREMENT

SETTINGS

,

page 73).

The values of the coefficients are provided with the sensor cap.

1.

Change the digit of the highlighted position with <><>.

2.

Go to the next position with <><>.

3.

Confirm with<ENTER> when a coefficient is completely entered.

Menu item

Possible
setting

Explanation