13.3 Conductivity

Error message,

OFL

Error message,

Error

– Measured temperature value outside

the operating conditions
(display of OFL/UFL instead of a
temperature value)

– Keep to the temperature range

for the test sample

– Defective sensor

– Calibrate
– Change sensor cap
– Replace sensor

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

Cause

Remedy

– Measured value outside the measur-

ing range

– Use suitable IDS conductivity

sensor

Cause

Remedy

– Sensor contaminated

– Clean the sensor and replace it

if necessary

– Calibration solution not suitable

– Check the calibration solutions