6.2 OPERATING MODE 1 - Output adjusted

manually -

In this mode, the pump’s motor speed is adjusted
manually using the front panel touch pad. The motor
speed can be adjusted from 0-100%. To adjust the
speed:

4 Set the pump for mode 1. Press the MODE button

until MODE 1 is shown on the LCD display. The
%SPEED icon will light. The large 3-DIGIT LCD
will indicate the currently programmed percentage of
speed.

4 Enter the programming mode. At the same time,

press the RUN/STANDBY button and the MODE
button. A blinking ARROW will point to the word
PROGRAM indicating the program mode has been
activated.

4 Press the FIELD button to select the digit to program.

The digit will blink when selected.

4 Press the DIGIT button to change the selected digit.

4 Repeat until all digits are programmed.

4 To exit the programming mode, press the

RUN/STANDBY button and the MODE button at the
same time. The arrow next to the word PROGRAM will
disappear.

[ NOTE: If while in the program mode no buttons are

pressed within 20 seconds, the circuitry will automati-
cally return to the run mode, without saving changes

OPERATING MODE 2 - Output adjusted
by 4-20 mA input signal

In this mode, the pump’s motor speed is adjusted
automatically based on the value of the 4-20 mA input
signal. Any motor speed can be assigned to either the
minimum or maximum milliamp input values.
However, the programmed minimum mA value
must be less than the programmed maximum mA
value.
The ALARM and SERVICE icons will blink
if the programming is in error. To assign the minimum
and maximum motor speed and the minimum and
maximum mA input signal values:

4 Set the pump for mode 2. Press the MODE button

until MODE 2 is shown on the LCD display. The
%SPEED or mA icon will light depending on the
current display setting. The large 3-DIGIT LCD will
indicate the current motor speed or the current mA

6.3

-

Page 10

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

% SPEED

1

MODE

100

RUN MODE 1

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

% SPEED

1

MODE

000

PROGRAM MODE 1

constant speed % setting

RUN MODE 2

RUN

PROGRAM

STAND-BY

PRIME

MINIMUM

MAXIMUM

2

MODE

04.0

mA

A-100NV