LaMotte COD Plus Colorimeter 1922-EX2 User Manual

Page 106

Advertising
background image

PROCEDURE

1.

Press and hold

ON

button until colorimeter turns on.

2.

Press

ENTER

to start.

3.

Press

ENTER

to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing 20 Phosphate L)
from TESTING MENU.

5.

Scroll to and select 20 Phosphate L from menu.

6.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

7.

Insert tube into chamber, close lid and select SCAN BLANK.

8.

Remove tube from colorimeter. Use 1.0 mL pipet (0354) to add 1.0 mL of
*Phosphate Acid Reagent (V-6282). Cap and mix.

9.

Use the 0.1 g spoon (0699) to add one measure of *Phosphate Reducing
Reagent (V-6283). Cap and shake until powder dissolves. Wait 5 minutes
for full color development. Solution will turn blue if phosphates are
present.

10. At end of 5 minute waiting period, mix, insert tube into chamber, close lid

and select SCAN SAMPLE. Record result.

11. Press

OFF button to turn colorimeter off or press EXIT button to exit to a

previous menu or make another menu selection.

COD PLUS TEST PROCEDURES 6/02

PHOSPHATE - LOW RANGE 3/3

Advertising
This manual is related to the following products: