ENTERING A NEW CALIBRATION

To enter a new calibration two reacted standards solutions of known
concentration are required: a “low standard” and a “high standard”. These
should be ready to use.

Use the

t button to scroll to New Calibration

and press

Q/ENTER button to select.

EDIT USER TEST01

* Select Vial/WL

New Calibration

Decimal Places

Select Units

Input the concentration of the LOW STANDARD by
using the

t and s buttons to scroll the first digit of

the concentration into the first position on the
display. Press

Q/ENTER button to select that digit

(1 for this example).

LOW STANDARD

0______

* Continue

The number “0” is always the starting point for the
next digit. Continue selecting digits or a decimal
point to enter the concentration (up to seven
characters).

LOW STANDARD

10_____

* Continue

“1.5” has been entered in this example. Press

Q/ENTER button four times to input “0” as the last

four digits. Pressing

Q/ENTER after selecting the

last digit saves the concentration.

LOW STANDARD

1.50___

* Continue

Input the concentration of the HIGH STANDARD by
using the same method as for the low standard.

HIGH STANDARD

0______

* Continue