PROCEDURE - SALT WATER

1.

Press and hold

ON

button until colorimeter turns on.

2.

Press

ENTER

to start.

3.

Press

ENTER

to select TESTING MENU.

4.

Select ALL TESTS (or another sequence containing
5 Ammonia-NLS

) from TESTING MENU.

5.

Scroll to and select 5 Ammonia-NLS from menu.

6.

Rinse a clean tube (0290) with sample water. Fill to the 10 mL line with
sample.

7.

Insert tube into chamber, close lid and select SCAN BLANK

.(See Note.)

8.

Remove tube from colorimeter. Use the 1.0 mL plastic pipet (0354) to add
2.0 mL of *Salicylate Ammonia #1 (3978). Cap and mix.

9.

Use the 0.15 g spoon (0727) to add two measures of *Salicylate #2
Reagent (7457). Cap and mix until dissolved. Wait 1 minute.

10. At end of 1 minute waiting period use 0.1 g spoon (0699) to add two

measures of *Salicylate #3 Reagent Powder (7458). Cap and shake
vigorously for at least 30 seconds and all solid has dissolved. Wait 20
minutes for maximum color development.

11. At the end of the 20 minute waiting period, immediately mix and insert

tube into chamber, close lid and select SCAN SAMPLE. Record result.

12. Press

OFF

button to turn colorimeter off or press

EXIT button to exit to a

previous menu or make another menu selection.

CALCULATIONS:
To express results as Unionized Ammonia (NH

3

):

ppm Unionized Ammonia (NH

3

) =

ppm Ammonia-Nitrogen (NH

3

–N) x 1.2

To express results as Ionized Ammonia (NH

4

):

ppm Ionized Ammonia (NH

4

+

) =

ppm Ammonia-Nitrogen (NH

3

–N) x 1.3

To determine the percentages of Unionized and Ionized Ammonia-Nitrogen,
consult the Appendix.
NOTE: For the best possible results, a reagent blank should be determined to
account for any contribution to the test result by the reagent system. To
determine the reagent blank, follow the above test procedure to scan a distilled
or deionized water blank. Then follow the above procedure to perform the test
on a distilled or deionized water sample. This test result is the reagent blank.
Subtract the reagent blank from all subsequent test results of unknown samples.
It is necessary to determine the reagent blank only when a new lot number of
reagents are obtained.

COD PLUS TEST PROCEDURES 2/04

AMMONIA-NITROGEN - LOW RANGE 1/3