Blank done, Storing – LaMotte COD Plus Colorimeter 1922-EX2 User Manual
Page 43
Advertising

Place a clear blank in the sample chamber. Press the
Q/ENTER button to scan the blank.
Insert Blank
* Continue
The screen will display Blank Done for about 1
second.
Blank Done
* Scan Blank
Place the reacted low standard in the sample
chamber. Press
Q/ENTER to scan the low standard.
Insert Lo Standard
* Continue
Place the reacted high standard in the sample
chamber. Press
Q/ENTER to scan the high
standard.
Insert Hi Standard
* Continue
The meter will display the message “Storing” and
return to the EDIT USER TEST01 menu.
Storing
EDIT USER TEST01
* New Calibration
Decimal Places
Select Units
COD PLUS COLORIMETER 6/02
43