The COD PLUS Colorimeter is ready to scan at the
correct wavelength. Place the blank in the sample
chamber, close the lid and press the

Q/ENTER

button to scan blank.
NOTE: Do not keep the button depressed.

001 COD LR

* Scan Blank

The screen will display Blank Done for about 1
second. Scan Sample will be positioned next to *.

001 COD LR

Blank Done

* Scan Blank

Place the reacted sample in the chamber, close the
lid and press the

Q/ENTER button to scan sample.

The colorimeter will scan the sample and the results
screen will appear.

001 COD LR

* Scan Sample

Record test result. To repeat the test, press the

Q/ENTER button to scan the sample again. The

last blank scanned is used to zero the colorimeter for
repeated scans. A different blank can be used by
pressing the

s button to scroll back to Scan

Blank

and then scanning another blank. Scroll with

the

t or s buttons and make another selection

with the

Q/ENTER button. The %T or Absorbance

of the last test can be viewed by choosing %T/Abs.
Press the

EXIT button to escape to previous menus.

NOTE: The menus loop in this screen so either the

s or t buttons will lead to the menu selection

needed.

001 COD LR

100 mgL

* Scan Blank

Next Test

Previous Test

%/Abs

Scan Blank

COD PLUS COLORIMETER 6/02