Conductivity

ProLab 4000

142

ba75646d03

12/2012

10.2.3

Measurement settings

The following measurement settings are possible for conductivity mea-
suring cells:

Menu item

Setting

Description

Measuring

– Cell constant

Calibrated cell constant

Measuring cells whose cell constant
is determined by calibration in KCL
control standard (calibration ranges
see page 157).
The currently valid cell constant is
displayed in the status line.

Adjustable cell constant...

Any measuring cells with freely
adjustable cell constants in the range
0.250 to 25.000 cm

-1

.

Adjustable LF 313T...

Measuring cell LF313T(-ID), nominal
cell constant 0.100 cm

-1

.

The cell constant can be adjusted in
the range 0.090 to 0.110 cm

-1

.

LF 213T

Measuring cell with a nominal cell
constant of 0.010 cm

-1

.

The cell constant is permanently set.

Temperature coefficient

TC nLF

Nonlinear temperature compensa-
tion

TC Lin...

0.000 ... 3.000 %/K

Coefficient of the linear temperature
compensation.

TC off

No temperature compensation.

TC nLin1......TC nLin4...

Measurement with entered or deter-
mined temperature coefficient.