1 determining the cell constant (calibration) – Xylem ProLab 4000 User Manual
Page 159

ProLab 4000
Conductivity
159
ba75646d03
12/2012
10.3.1
Determining the cell constant (calibration)
Proceed as follows to determine the cell constant:
1
Connect a conductivity measuring cell to the meter.
2
Select the channel of the measuring cell in the measured value
display.
3
If necessary, select the parameter, conductivity with <MODE>.
4
Start the calibration with <CAL>.
The calibration display appears.
5
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
6
Start the measurement with [OK].
The measured value is checked for stability (Stability control).
7
Wait for the end of the measurement with Stability control or
accept the calibration value with [OK].
The cell constant determined is displayed.
8
Accept the new calibration values with [OK].
The calibration record is displayed and output to the interface.
The calibration procedure is finished.
or
To switch to the measured value display without accepting the
new calibration values, press <ESC>.