Calibration points
Calibration can be performed using one to five buffer solutions in any
order (single-point to five-point calibration). The measuring instrument
determines the following values and calculates the calibration line as
follows:
19
Reagecon TEC *
2,000
4,010
7,000
10,000
25 °C
20
Reagecon 20 *
2,000
4,000
7,000
10,000
13,000
20°C
21
Reagecon 25 *
2,000
4,000
7,000
10,000
13,000
25 °C
22
Riedel-de Haen *
2,000
4,000
7,000
10,000
20°C
No.
Buffer set *
pH values
at
Note
The buffers are selected in the Calibration sensor menu (see
page 73).
Displayed calibration data
1-point
Zero point
Slope = Nernst slope (-59.16 mV/pH at 25 °C)
2-point
Zero point
Slope
3-point to
5-point
Zero point
Slope
The calibration line is calculated by linear regression.