Xylem PH3110 FIELD PH METER User Manual

Page 26

Advertising
background image

Operation

pH 3110

26

ba75791e02

11/2008

Note
While the zero point (ASY) is being displayed, you can change the unit
of the zero point with

<

▲><▼>.

While the slope (SLO) is being displayed, you can change the unit of
the slope with

<

▲><▼>.

The % display refers to the Nernst slope of 59.2 mV/pH at 25° C
(100 x determined slope/Nernst slope).

The units of the zero point and slope can also be changed permanently
in the measurement settings (see section 4.5.2).

7

To continue the two-point calibration, thoroughly rinse the
combination electrode with deionized water.

8

If necessary, set the temperature of the second buffer solution
with <

▲><▼>.

9

Immerse the combination electrode in the second buffer
solution.

10

Press the <ENTER> key. The second pH value appears on the
display.

11

When the measured value is stable,
using <

▲><▼>, set the measured value to the nominal pH

value of the buffer solution (at the current temperature).

12

Press the <ENTER> key.
The value of the zero point (ASY) is displayed.

13

Press the <ENTER> key.
The value of the slope (SLO) is displayed.

14

Press the <ENTER> key.
The measuring mode is active.