6 reset, 1 resetting calibration values, Reset 4.6.1 – Xylem PH3110 FIELD PH METER User Manual
Page 34: Resetting calibration values
Advertising

Operation
pH 3110
34
ba75791e02
11/2008
4.6
Reset
4.6.1
Resetting calibration values
This function resets the calibration values to the default condition. All
other meter settings are retained.
Calibration values in the
default condition
Note
The measuring system is not calibrated after a reset. Before
measuring, recalibrate the meter.
Resetting calibration
values
Zero point
0 mV (pH 7.000)
Slope
-59.16 mV/pH (100 %)
1
Press <On/Off
_
> to open the menu for the reset of the
calibration data.
Init.C is displayed.
2
Press <
▲><▼> to display no or YES.
YES: Reset the calibration values.
no: Retain the calibration values.
3
Confirm with <ENTER>.
The menu is finished.
The meter switches to the measuring mode.