Xylem PH3110 FIELD PH METER User Manual
Page 41
Advertising

pH 3110
What to do if...
41
ba75791e02
11/2008
Obviously incorrect
measured values
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause
Remedy
pH combination electrode:
– pH combination electrode
unsuitable
– Use suitable combination
electrode
– Temperature difference
between buffer and test sample
too high
– Adjust temperature of buffer
or sample solutions
– Measurement procedure not
suitable
– Follow special procedure
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <ENTER> and
<On/Off>
key
simultaneously
Cause
Remedy
– E. g., a question by the service
department
– Switch on the meter.
During the display test,
display the software version
with <ENTER>.