6 what to do if, What to do if, Chapter 6 w – Xylem PH3110 FIELD PH METER User Manual
Page 39: 6what to do if

pH 3110
What to do if...
39
ba75791e02
11/2008
6
What to do if...
Error message
OFL, UFL
Error message
CalError
Cause
Remedy
pH combination electrode:
– Measured value outside the
measuring range
– Use suitable combination
electrode
– Air bubble in front of the
diaphragm
– Remove air bubble
– Air in the diaphragm
– Extract air or moisten
diaphragm
– Cable broken
– Replace combination
electrode
– Gel electrolyte dried out
– Replace combination
electrode
Cause
Remedy
pH combination electrode:
– The values determined for zero
point and slope of the
combination electrode are
outside the allowed limits.
– Recalibrate
– Diaphragm contaminated
– Clean diaphragm
– Combination electrode broken
– Replace combination
electrode
Buffer solutions
– Incorrect buffer solutions
– Change calibration
procedure
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions