2 measurement settings, Measurement settings, Fu b – Xylem PH3110 FIELD PH METER User Manual
Page 30

Operation
pH 3110
30
ba75791e02
11/2008
4.5.2
Measurement settings
These settings apply to calibration and measurement (the default
setting is printed in bold).
Display during
calibration
(bUFF)
Display during calibration (BUFFER)
pH (buffer nominal value),
U (combination electrode
voltage)
Unit of the value for the slope (SLO)
mV/pH, %
Unit of the value for the zero point
(ASY)
mV, pH
Measured value resolution, pH
0.000 , 0.00 , 0.0
Measured value resolution, U
0.0 , 0
Temperature unit (UnIt)
°C, °F
Calibration interval (Int.C)
0 ... 7 ... 999 d
1
With <M
_
> open the menu for measurement and calibration
settings.
The first setting is displayed.
2
Using <
▲><▼>, select the pH or U display during calibration.
3
Confirm with <ENTER>.
SLO, the unit of the value for the slope (mV/pH or %) is
displayed.
H
p
F
F
U
b
pH
AutoCal DIN
AutoCal TEC