Malfunctions: Causes and elimination

8

Malfunctions: Causes and elimination

Fault

Cause

Troubleshooting

Important

Machine is switched
off by the motor pro-
tection switch

Mains voltage/ Frequency
does not correspond with the
motor data

Check by qualifi ed electrician

Section 5.5

Connection to motor terminal
board is not correct

Motor protection switch is not
set correctly

Motor protection switch is
triggered too quickly

Use a motor protection switch
with an overload-dependent
delayed switch off that takes
into consideration the short
term excess current at start
up (version with short circuit
and overload trigger as per
VDE 0660 Part 2 orIEC 947-4)

Pumping capacity is
insuffi cient

The mesh fi lter is dirty

Clean or replace the mesh
fi lter

Section 7.2.2
Section 7.4

The suction pipe is too long or
too narrow

Check the hose or the pipe

Section 5.3

Machine or system leaking

Check the pipework and
screw connections for leaks
and to ensure that they are
fi rmly seated

Section 7.2