Sourcemeter, Line, Series 2400 – Atec Keithley-2400 User Manual

Page 2

Advertising
background image

www.keithley.com

1.888.KEITHLEY

(U.S. only)

S

O

UR

C

E

A

ND

M

E

A

S

UR

E

Ti

gh

tly

c

oup

le

d

pre

ci

si

on

s

our

ci

ng

a

nd

me

as

ure

me

nt

Ordering Info

2400

200V, 1A, 20W

SourceMeter

2400-C

200V, 1A, 20W

SourceMeter with

Contact Check

2400-LV 20V, 1A, 20W

SourceMeter

2410

1100V, 1A, 20W

SourceMeter

2410-C

1100V, 1A, 20W

SourceMeter with

Contact Check

2420

60V, 3A, 60W

SourceMeter

2420-C

60V, 3A, 60W

SourceMeter with

Contact Check

2425

100V, 3A, 100W

SourceMeter

2425-C

100V, 3A, 100W

SourceMeter with

Contact Check

2430

100V, 10A, 1000W Pulse

Mode SourceMeter

2430-C

100V, 10A, 1000W Pulse

Mode SourceMeter

with Contact Check

2440

40V, 5A, 50W

SourceMeter

2440-C

40V, 5A, 50W

SourceMeter with

Contact Check

Accessories Supplied

Test Leads
LabVIEW Software Driver

(downloadable)
LabTracer Software

(downloadable)
TestPoint Software Driver

(downloadable)

Local

Remote

Remote

Local

DUT

IN/OUT HI

SENSE HI

SENSE LO

IN/OUT LO

I

meter

I

source

V

meter

/Compliance

Local

Remote

Remote

Local

DUT

IN/OUT HI

SENSE HI

SENSE LO

IN/OUT LO

I

meter

/Compliance

V

source

V

meter

Feedback to

Adjust V

source

Source I, Measure V, I, or

Ω configuration

Source V, Measure I, V, or

Ω configuration

+1A

–1A

+100mA

+20V

–20V

+200V

–200V

–100mA

2400

only

2400

only

Duty cycle

limited

Model 2400 and 2400-LV SourceMeter

+1A

–1A

+200mA

+20V

–20V

+200V

–200V

–100mA

+20mA

–20mA

+1100V

–1100V

Duty cycle

limited

Model 2410 High-Voltage SourceMeter

+1A

+20V

–20V

–1A

+100mA

–100mA

+3A

–60V

+60V

–3A

Duty cycle

limited

Model 2420 3A SourceMeter

+1A

+60

–100mA

+20

–20

–60

+100V

–100V

+3A

+100mA

–1A

–3A

Duty cycle

limited

Model 2425 100W SourceMeter

+1A

+60

–100mA

+20

–20

–60

+100V

+3A

+100mA

+10A

–1A

–3A

–10A

Pulse mode

only

Duty cycle

limited

–100V

Model 2430 1kW Pulse Mode SourceMeter

5A

3A

100mA

–10V

+10V

–100mA

-–3A

-5A

–40V

40V

1A

–1A

Duty cycle

limited

Model 2440 5A SourceMeter

I-V Characteristics
All SourceMeter instruments provide four-quadrant operation. In the first and third quadrants they

operate as a source, delivering power to a load. In the second and fourth quadrants they operate as a

sink, dissipating power internally. Voltage, current, and resistance can be measured during source or

sink operation.