Sourcemeter, Line, Series 2400 – Atec Keithley-2400 User Manual

Page 8: Se rie s 24 00 co nd ens ed spe ci fic at io ns, System speeds

Advertising
background image

www.keithley.com

1.888.KEITHLEY

(U.S. only)

S

O

UR

C

E

A

ND

M

E

A

S

UR

E

Se

rie

s

24

00

co

nd

ens

ed

spe

ci

fic

at

io

ns

GENERAL

System Speeds

MEASUREMENT

1

MAXIMUM RANGE CHANGE RATE: 75/second.
MAXIMUM MEASURE AUTORANGE TIME: 40ms (fixed source).

2

Sweep Operation

3

Reading Rates (rdg./second) for 60Hz (50Hz):

Measure

Source-Measure

Source-Measure

5

Pass/Fail Test

4, 5

Source-Memory

4

Speed

NPLC/Trigger Origin

To Mem.

To GPIB

To Mem.

To GPIB

To Mem.

To GPIB

To Mem.

To GPIB

Fast

0.01 / internal

2081 (2030)

1754

1551 (1515)

1369

902 (900)

981

165 (162)

165

IEEE-488.1 Mode

0.01 / external

1239 (1200)

1254

1018 (990)

1035

830 (830)

886

163 (160)

163

Fast

0.01 / internal

2081 (2030)

1198 (1210)

1551 (1515)

1000 (900)

902 (900)

809 (840)

165 (162)

164 (162)

IEEE-488.2 Mode

0.01 / external

1239 (1200)

1079 (1050)

1018 (990)

916 (835)

830 (830)

756 (780)

163 (160)

162 (160)

Medium

0.10 / internal

510

(433)

509 (433)

470 (405)

470 (410)

389 (343)

388 (343)

133 (126)

132 (126)

IEEE-488.2 Mode

0.10 / external

438

(380)

438 (380)

409 (360)

409 (365)

374 (333)

374 (333)

131 (125)

131 (125)

Normal

1.00 / internal

59

(49)

59

(49)

58

(48)

58

(48)

56

(47)

56

(47)

44

(38)

44

(38)

IEEE-488.2 Mode

1.00 / external

57

(48)

57

(48)

57

(48)

57

(47)

56

(47)

56

(47)

44

(38)

44

(38)

Single Reading Operation Reading Rates (rdg./second) for 60Hz (50Hz):

Speed

NPLC/Trigger Origin

Measure

To GPIB

Source-Measure

5

To GPIB

Source-Measure Pass/Fail Test

4,5

To GPIB

Fast (488.1)

0.01 / internal

537

140

135

Fast (488.2)

0.01 / internal

256 (256)

79 (83)

79 (83)

Medium (488.2)

0.10 / internal

167

(166)

72 (70)

69 (70)

Normal (488.2)

1.00 / internal

49

(42)

34

(31)

35 (30)

Component for 60Hz (50Hz):

4, 6

Speed

NPLC/Trigger Origin

Measure To GPIB

Source Pass/Fail Test

Source-Measure Pass/Fail Test

5, 7

To GPIB

Fast

0.01 / external

1.04 ms (1.08 ms)

0.5 ms (0.5 ms)

4.82 ms (5.3 ms)

Medium

0.10 / external

2.55 ms (2.9 ms)

0.5 ms (0.5 ms)

6.27 ms (7.1 ms)

Normal

1.00 / external

17.53 ms (20.9 ms)

0.5 ms (0.5 ms)

21.31 ms (25.0 ms)

1

Reading rates applicable for voltage or current measurements. Auto zero off, autorange off, filter off, display off,

trigger delay = 0, and binary reading format.

2

Purely resistive lead. 1µA and 10µA ranges <65ms.

3

1000 point sweep was characterized with the source on a fixed range.

4

Pass/Fail test performed using one high limit and one low math limit.

5

Includes time to re-program source to a new level before making measurement.

6

Time from falling edge of START OF TEST signal to falling edge of END OF TEST signal.

7

Command processing time of :SOURce:VOLTage|CURRent:TRIGgered <nrf> command not included.

Noise Rejection:

NPLC

NMRR

CMRR

Fast

0.01

80 dB

Medium

0.1

80 dB

Slow

1

60 dB

100 dB

1

1

Except lowest 2 current ranges = 90dB.

LOAD IMPEDANCE: Stable into 20,000pF typical.
COMMON MODE VOLTAGE: 250V DC (40V DC for Model 2440).
COMMON MODE ISOLATION: >10

9

Ω, <1000pF.

OVERRANGE: 105% of range, source and measure.
MAX. VOLTAGE DROP BETWEEN INPUT/OUTPUT AND SENSE TERMINALS: 5V.
MAX. SENSE LEAD RESISTANCE: 1M

Ω for rated accuracy.

SENSE INPUT IMPEDANCE: >10

10

Ω.

GUARD OFFSET VOLTAGE: <150µV, typical (300µV for Models 2430, 2440).
SOURCE OUTPUT MODES:

Pulse (Model 2430 only)
Fixed DC level
Memory List (mixed function)
Stair (linear and log)

MEMORY BUFFER: 5,000 readings @ 5 digits (two 2,500 point buffers). Includes selected measured

value(s) and time stamp. Lithium battery backup (3 yr+ battery life).

SOURCE MEMORY LIST: 100 points max.
PROGRAMMABILITY: IEEE-488 (SCPI-1995.0), RS-232, 5 user-definable power-up states plus

factory default and *RST.

DIGITAL INTERFACE:
Interlock: Active low input.
Handler Interface: Start of test, end of test, 3 category bits. +5V@ 300mA supply.
Digital I/O: 1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA, diode clamped).
POWER SUPPLY: 100V to 240V rms, 50–60Hz (automatically detected at power up). Model

2400: 190VA. Model 2410: 210VA. Model 2420: 220VA. Model 2425, 2430: 250VA. Model
2440:
240VA.

COOLING: (Model 2410, 2420, 2425, 2430, 2440): Forced air, variable speed.
EMC: Conforms to European Union Directive 89/336/EEC, EN 61326-1.
SAFETY: Conforms to European Union Directive 73/23/EEC, EN61010-1.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARM-UP: 1 hour to rated accuracies.
DIMENSIONS: 89mm high

× 213mm wide × 370mm deep (3

1

2

in

× 8

3

8

in

× 14

9

16

in). Bench

Configuration (with handle & feet):104mm high

× 238mm wide × 370mm deep (4

1

8

in

×

9

3

8

in

× 14

9

16

in).

WEIGHT: 3.21kg (7.08 lbs) (Model 2425, 2430, 2440: 4.1kg, 9.0 lbs).
ENVIRONMENT: Operating: 0°–50°C, 70%R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.

Storage: –25°C to 65°C.

Series 2400

SourceMeter

®

Line

Advertising