Digital I/O Interface
The digital I/O interface can link the SourceMeter instrument to many

popular component handlers, including Aetrium, Aeco, and Robotronics.

Other capabilities of the interface include:
• Tight systems integration for applications such as binning and sorting
• Built-in component handler interface
• Start of test and end of test signals
• 5V, 300mA power supply
• Optional expander accessory (Model 2499-DIGIO) adds 16 digital I/O lines

Trigger Link Interface
All SourceMeter instruments include Keithley’s unique Trigger Link inter-

face which provides high-speed, seamless communications with many of

Keithley’s other instruments. For example, use the Trigger Link interface

to connect a SourceMeter instrument with a Series 7000 Switching System

for a complete multi-point test solution. With Trigger Link, the 7000 Series

Switching Systems can be controlled by a SourceMeter during a high-speed

test sequence independent of a computer and GPIB.

Optional Contact Check Function
The Contact Check function makes it simple to verify good connections

quickly and easily before an automated test sequence begins. This elimi-

nates measurement errors and false product failures associated with con-

tact fatigue, breakage, contamination, loose or broken connection, relay

failures, etc. Some capabilities of this function are:
• 350µs verification and notification process time
• The output of the SourceMeter instrument is automatically shut off after

a fault and is not re-activated until good contact is verified, protecting

the device under test from damage and the operator from potential

safety hazards

• 3 pass/fail threshold values: 2

Ω, 15Ω, and 50Ω

• No energy passes through the device under test during the operation
• Enabled either from the front panel or remotely over the GPIB
• 3 fault notification methods

Unique 6-Wire Ohms Technique
SourceMeter instruments can make standard 4-wire, split Kelvin, and

6-wire, guarded ohms measurements and can be configured for either the

constant current or constant voltage method. The 6-wire ohms technique:
• Uses guard and guard sense leads in addition to the 4-wire sense and

source leads

• Locks out parallel current paths when measuring resistor networks or

hybrid circuits to isolate the component under test

• Allows users to configure and plot data easily from Series 2400

SourceMeter instruments, making characterization of two, three, and

four terminal devices a snap

+

-

I

meter

V

meter

V or I

Source

350

µs

Contact Check

(optional)

GUARD

GUARD SENSE

IN/OUT HI

SENSE HI

SENSE LO

IN/OUT LO

Pass

Pass

Fail

Contact check option for 4-wire or 6-wire applications

Free LabTracer 2.0 device characterization software (downloadable)

+

-

GUARD

GUARD SENSE

IN/OUT HI

SENSE HI

SENSE LO

IN/OUT LO

DUT

R2

R1

R3

V

meter

I

meter

I

source

6-Wire Ohms Circuit. All test current flows through R1 because the

high current guard drives the voltage across R2 to 0V.

Series 2400

SourceMeter

®

Line