Voltage Accuracy (Local or Remote Sense)

Model

Range

Programming

Resolution

Source

1

Accuracy (1 Year)

23°C ±5°C

±(% rdg. + volts)

Default

Measurement

Resolution

Measurement

2, 3, 4

Accuracy (1 Year)

23°C ±5°C

±(% rdg. + volts)

Output Slew Rate

(±30%)

Source/Sink Limit

2400, 2400-C

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±21 V @ ±1.05 A

±210 V @ ±105 mA

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

20.0000 V

500 µV

0.02% + 2.4 mV

100 µV

0.015% + 1.5 mV

0.08 V/µs

200.000 V

5 mV

0.02% + 24 mV

1 mV

0.015% + 10 mV

0.5 V/µs

2400-LV

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±21 V @ ±1.05 A

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

20.0000 V

500 µV

0.02% + 2.4 mV

100 µV

0.015% + 1.5 mV

0.08 V/µs

2410, 2410-C

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±21 V @ ±1.05 A

±1100 V @ ±21 mA

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

20.0000 V

500 µV

0.02% + 2.4 mV

100 µV

0.015% + 1 mV

0.15 V/µs

1000.00 V

50 mV

0.02% + 100 mV

10 mV

0.015% + 50 mV

0.5 V/µs

2420, 2420-C

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±21 V @ ±3.15 A
±63 V @ ±1.05 A

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

20.0000 V

500 µV

0.02% + 2.4 mV

100 µV

0.015% + 1 mV

0.08 V/µs

60.0000 V

1.5 mV

0.02% + 7.2 mV

1 mV

0.015% + 3 mV

0.14 V/µs

2425, 2425-C

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±21 V @ ±3.15 A

±105 V @ ±1.05 A

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

20.0000 V

500 µV

0.02% + 2.4 mV

100 µV

0.015% + 1 mV

0.08 V/µs

100.0000 V

2.5 mV

0.02% + 12 mV

1 mV

0.015% + 5 mV

0.25 V/µs

2430, 2430-C

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±105 V @ ±1.05 A

±105 V @ ±10.5 A
(pulse mode only)

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

20.0000 V

500 µV

0.02% + 2.4 mV

100 µV

0.015% + 1 mV

0.08 V/µs

100.0000 V

2.5 mV

0.02% + 12 mV

1 mV

0.015% + 5 mV

0.25 V/µs

2440, 2440-C

200.000 mV

5 µV

0.02% + 600 µV

1 µV

0.012% + 300 µV

±10.5 V @ ±5.25 A

±42 V @ ±1.05 A

2.00000 V

50 µV

0.02% + 600 µV

10 µV

0.012% + 300 µV

10.0000 V

500 µV

0.02% + 1.2 mV

100 µV

0.015% + 750 µV

0.08 V/µs

40.0000 V

5 mV

0.02% + 4.8 mV

1 mV

0.015% + 3 mV

0.25 V/µs

TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15

× accuracy specification)/°C.

VOLTAGE REGULATION: Line: 0.01% of range. Load: 0.01% of range + 100µV.
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = none.
CURRENT LIMIT: Bipolar current limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT: <0.1% typical (full scale step, resistive load, 10mA range).

ADDITIONAL SOURCE SPECIFICATIONS (All Models)

TRANSIENT RESPONSE TIME: 30µs minimum for the output to recover to its spec. following a

step change in load.

COMMAND PROCESSING TIME: Maximum time required for the output to begin to change

following the receipt of :SOURce:VOLTage|CURRent <nrf> command. Autorange On: 10ms.
Autorange Off: 7ms.

OUTPUT SETTLING TIME: Time required to reach 0.1% of final value after command is pro-

cessed. 100µs typical. Resistive load. 10µA to 100mA range.

DC FLOATING VOLTAGE: Output can be floated up to ±250VDC (Model 2440 ±40VDC) from

chassis ground.

REMOTE SENSE: Up to 1V drop per load lead.
COMPLIANCE ACCURACY: Add 0.3% of range and ±0.02% of reading to base specification.
OVER TEMPERATURE PROTECTION: Internally sensed temperature overload puts unit in

standby mode.

RANGE CHANGE OVERSHOOT: Overshoot into a fully resistive 100k

Ω load, 10Hz to 1MHz BW,

adjacent ranges: 100mV typical, except 20V/200V (20V/60V on Model 2420), 20V/100V on Model
2425 and 2430, range boundary, and Model 2440.

MINIMUM COMPLIANCE VALUE: 0.1% of range.

ADDITIONAL PULSE MODE SOURCE SPECIFICATIONS

(2430 and 2430-C only)

MAXIMUM DUTY CYCLE: 8%, hardware limited, 10A range only. All other ranges 100%.
MAXIMUM PULSE WIDTH: 5ms from 90% rising to 90% falling edge, 2.5ms 10A range.
MINIMUM PULSE WIDTH: 150µs.
MINIMUM PULSE RESOLUTION: 50µs typical, 70µs max., limited by system jitter.
SOURCE ACCURACY: Determined by settling time and source range specifications.
OUTPUT SETTLING TIME 0.1%:

800µs typ., source I = 10A into 10

Ω, limited by voltage slew rate.

500µs typ., source I = 10A into 1

Ω, limited by voltage slew rate.

OUTPUT SLEW RATE:

Voltage (10

load): 0.25V/µs ±30% on 100V range. 0.08V/µs ±30% on 20V range, 10A range.

Current (0

load): 0.25A/µs ±30% on 100V range. 0.08A/µs ±30% on 20V range, 10A range.

NOTES

1. 2400, 2410 Only: Specifications valid for continuous output currents below 105mA. For operation above 105mA

continuous for >1 minute, derate accuracy 10%/35mA above 105mA.

2. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A

ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A ranges, add

0.5%.

3. Accuracies apply to 2- or 4-wire mode when properly zeroed.
4. In pulse mode, limited to 0.1 PLC measurement.

Series 2400

SourceMeter

®