Sourcemeter, Line, Series 2400 – Atec Keithley-2400 User Manual

Page 6: Se rie s 24 00 co nd ens ed spe ci fic at io ns, Current accuracy (local or remote sense)

Advertising
background image

www.keithley.com

1.888.KEITHLEY

(U.S. only)

S

O

UR

C

E

A

ND

M

E

A

S

UR

E

Se

rie

s

24

00

co

nd

ens

ed

spe

ci

fic

at

io

ns

Current Accuracy (Local or Remote Sense)

Model

Range

Programming

Resolution

Source

1, 3

Accuracy

(1 Year)

3

23°C ±5°C

±(% rdg. + amps)

Default Measurement

Resolution

Measurement

5, 6, 7

Accuracy (1 Year)

23°C ±5°C

±(% rdg. + amps)

Source/Sink Limit

2400, 2400-C,
2400-LV

1.00000 µA

50 pA

0.035% + 600 pA

10 pA

0.029% + 300 pA

±1.05A @ ±21 V

±105 mA @ ±210 V

8

10.0000 µA

500 pA

0.033% + 2 nA

100 pA

0.027% + 700 pA

100.000 µA

5 nA

0.031% + 20 nA

1 nA

0.025% + 6 nA

1.00000 mA

50 nA

0.034% + 200 nA

10 nA

0.027% + 60 nA

10.0000 mA

500 nA

0.045% + 2 µA

100 nA

0.035% + 600 nA

100.000 mA

5 µA

0.066% + 20 µA

1 µA

0.055% + 6 µA

1.00000 A

2

50 µA

0.27 % + 900 µA

10 µA

0.22 % + 570 µA

2410, 2410-C

1.00000 µA

50 pA

0.035% + 600 pA

10 pA

0.029% + 300 pA

±1.05A @ ±21 V

±21 mA @ ±1100 V

10.0000 µA

500 pA

0.033% + 2 nA

100 pA

0.027% + 700 pA

100.000 µA

5 nA

0.031% + 20 nA

1 nA

0.025% + 6 nA

1.00000 mA

50 nA

0.034% + 200 nA

10 nA

0.027% + 60 nA

20.0000 mA

500 nA

0.045% + 4 µA

100 nA

0.035% + 1.2 µA

100.000 mA

5 µA

0.066% + 20 µA

1 µA

0.055% + 6 µA

1.00000 A

2

50 µA

0.27 % + 900 µA

10 µA

0.22 % + 570 µA

2420, 2420-C

10.0000 µA

500 pA

0.033% + 2 nA

100 pA

0.027% + 700 pA

±3.15A @ ±21 V

±1.05 A @ ±63 V

100.000 µA

5 nA

0.031% + 20 nA

1 nA

0.025% + 6 nA

1.00000 mA

50 nA

0.034% + 200 nA

10 nA

0.027% + 60 nA

10.0000 mA

500 nA

0.045% + 2 µA

100 nA

0.035% + 600 nA

100.000 mA

5 µA

0.066% + 20 µA

1 µA

0.055% + 6 µA

1.00000 A

2

50 µA

0.067% + 900 µA

10 µA

0.066% + 570 µA

3.00000 A

2

50 µA

0.059% + 2.7 mA

10 µA

0.052% + 1.71 mA

2425, 2425-C

10.0000 µA

500 pA

0.033% + 2 nA

100 pA

0.027% + 700 pA

±3.15A @ ±21 V

±1.05 A @ ±105 V

100.000 µA

5 nA

0.031% + 20 nA

1 nA

0.025% + 6 nA

1.00000 mA

50 nA

0.034% + 200 nA

10 nA

0.027% + 60 nA

10.0000 mA

500 nA

0.045% + 2 µA

100 nA

0.035% + 600 nA

100.000 mA

5 µA

0.066% + 20 µA

1 µA

0.055% + 6 µA

1.00000 A

2

50 µA

0.067% + 900 µA

10 µA

0.060% + 570 µA

3.00000 A

2

50 µA

0.059% + 2.8 mA

10 µA

0.052% + 1.71 mA

2430, 2430-C

10.0000 µA

500 pA

0.033% + 2 nA

100 pA

0.027% + 700 pA

±1.05A @ ±105 V

±10.5 A @ ±105 V
(pulse mode only)

100.000 µA

5 nA

0.031% + 20 nA

1 nA

0.025% + 6 nA

1.00000 mA

50 nA

0.034% + 200 nA

10 nA

0.027% + 60 nA

10.0000 mA

500 nA

0.045% + 2 µA

100 nA

0.035% + 600 nA

100.000 mA

5 µA

0.066% + 20 µA

1 µA

0.055% + 6 µA

1.00000 A

50 µA

0.067% + 900 µA

10 µA

0.060% + 570 µA

3.00000 A

2

10.00000 A

4

500 µA
500 µA

0.059% + 2.8 mA

0.089% + 5.9 mA

10 µA
10 µA

0.052% + 1.71 mA
0.082% + 1.71 mA

2440, 2440-C

10.0000 µA

500 pA

0.033% + 2 nA

100 pA

0.027% + 700 pA

±5.25A @ ±10.5 V

±1.05 A @ ±42 V

100.000 µA

5 nA

0.031% + 20 nA

1 nA

0.025% + 6 nA

1.00000 mA

50 nA

0.034% + 200 nA

10 nA

0.027% + 60 nA

10.0000 mA

500 nA

0.045% + 2 µA

100 nA

0.035% + 600 nA

100.000 mA

5 µA

0.066% + 20 µA

1 µA

0.055% + 6 µA

1.00000 A

50 µA

0.067% + 900 µA

10 µA

0.060% + 570 µA

5.00000 A

50 µA

0.10 % + 5.4 mA

10 µA

0.10 % + 3.42 mA

TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15

× accuracy specification)/°C.

CURRENT REGULATION: Line: 0.01% of range. Load: 0.01% of range (except Model 2440

5A range 0.05%) + 100pA.

VOLTAGE LIMIT: Bipolar voltage limit (compliance) set with single value. Min. 0.1% of range.
OVERSHOOT: <0.1% typical (1mA step, RL = 10k

Ω, 20V range for Model 2400, 2410, 2420, 2425,

2430), (10V range for Model 2440).

CONTACT CHECK SPECIFICATIONS (requires -C version)

SPEED: 350µs for verification and notification.
CONTACT CHECK:

2

15

50

No contact check failure

<1.00

<13.5

<47.5

Always contact check failure

>3.00

>16.5

>52.5

NOTES

1. 2400, 2410 Only: Specifications valid for continuous output currents below 105mA. For operation above 105mA

continuous for >1 minute, derate accuracy 10%/35mA above 105mA.

2. Full operation (1A) regardless of load to 30°C (50°C for Model 2420 and 2440). Above 30°C (50°C for Model

2420 and 2440) ambient, derate 35mA/°C and prorate 35mA/

Ω load. 4-wire mode. For current sink operation

on 1A, 3A, or 5A ranges, maximum continuous power is limited to approximately 1/2 rated power or less,

depending on current, up to 30°C ambient. See power equations in the User’s Manual to calculate allowable

duty cycle for specific conditions.

3. For sink mode, 1µA to 100mA range, accuracy is:

Model 2400: ±(0.15% + offset*4). Models 2410, 2420, 2425, 2430, 2440: ±(0.5% + offset*3).

For 1A range, accuracy is:

Model 2400: ±(1.5% + offset*8). Models 2410, 2420, 2425, 2430, 2440: ±(1.5% + offset*3).

4. 10A range only in pulse mode. Limited to 2.5ms pulse width maximum. 10% duty cycle maximum.
5. Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A

ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A ranges, add

0.5%.

6. Accuracies apply to 2- or 4-wire mode when properly zeroed.
7. In pulse mode, limited to 0.1 PLC measurement.
8. Model 2400 and 2400-C only.

Series 2400