2-3

Section 2 Installation

Models SLA5810 and SLA5820

Installation and Operation Manual
X-PR-SLA5800-PC-eng
Part Number: 541B091AAG
July, 2010

2-8 In-Line Filter

It is recommended that an in-line filter be installed upstream from the
pressure controller to prevent the possibility of any foreign material entering
the sensor or control valve. The filtering element should be replaced
periodically or ultrasonically cleaned.

Table 2-1 Recommended Filter Size

2-9 Installation

Recommended installation procedures:
a.The Digital PC should be located in a clean,

dry atmosphere relatively free from shock and vibration.

b.Leave sufficient room for access to Self-zero function push-button.
c. Install in such a manner that permits easy removal if the instrument

requires servicing.

d.The Digital PC can be installed in any position.

Maximum Flow Rate

Recommended Filter

100 sccm

2 micron

500 sccm

2 micron

1 to 5 slpm

10 micron

10 to 30 slpm

40 micron

10 to 30 slpm

40 micron

30 slpm to 100 slpm

40 micron

> 100 slpm

Consult Factory