Analog i/o pin connections – Brooks Instrument SLA5810/20 User Manual

Page 19

Advertising
background image

2-5

Section 2 Installation

Models SLA5810 and SLA5820

Installation and Operation Manual
X-PR-SLA5800-PC-eng
Part Number: 541B091AAG
July, 2010

Figure 2-2 Common Electrical Hookups Voltage I/O Version

Figure 2-1 D-Connector Shielded Cable Hookup Diagram - Analog I/O Options.

Analog I/O Pin Connections:

PIN #

FUNCTION

1

Setpoint Command Input (-)

2

Signal, 0(1)-5 volt, Output (+)

3

Not Connected

4
5

Power Supply, +15 Vdc or +24 Vdc (+)

6
7
8
9

Power Supply, Common (-)

10

Signal, Common, Output, (-)

11

Reference, +5 Vdc, Output, (+)

12

Valve Override, Input

13

Calibration Select, Input

14

Chassis Ground

15

External Sensor, Input

Signal, 0(4)-20 mA, Output (+)

Not Connected

Setpoint, 0(4)-20 mA, Input (+)
Setpoint, 0(1)-5 volt, Input (+)

Notes:

1. The commons for the setpoint and signal output are not isolated from power supply common.
2. The cable shields should connect to chassis ground at one end only (Pressure Controller only).
3. Input impedence of setpoint input is >990K

. Maximum allowable input signal is 20 Vdc.

4. Minimum load resistance of voltage output is 2K

.

or +24 Vdc

Advertising